NOC Oper
ation
Chapter 6
Diagnostics and Troubleshooting
6.1
About this chapter
This chapter provides information on the status alarms associated with the NOC application.
Note: For information on diagnostics or troubleshooting for the Series 3000 platform, see the manual 
entitled 
Series 3000 MVD Transmitters and Controllers: Configuration and Use Manual
6.2
NOC status alarms
Table 6-1 lists all status alarms associated with the NOC application, with descriptions and suggested 
user actions.
For information on other Series 3000 status alarms, see the manual entitled Series 3000 MVD 
Transmitters and Controllers: Configuration and Use Manual
.
Table 6-1
Process alarms
Alarm 
number
Maintenance 
menu listing
Description
User actions
A136
Power Outage
The power was off for at least 30 seconds.
No action required.
A137 Measurements 
Paused
Continuous mode measurements are paused.
No action required. If desired, 
resume measurement.
A138
TBR Active
The drive gain has exceeded the configured TBR 
threshold and the configured TBR action has 
been implemented.
No action required. 
A139
Water Cut 
Overrange
The water cut measured by the water cut monitor 
is above the configured External Water Cut Limit.
No action required.
A140
TMR active
The drive gain from the TMR gas meter has 
exceeded the configured TMR threshold.
No action required.