Default Values and Ranges
Density
Density damping
1.28 sec
0.0 – 60.0 sec
User-entered value is 
corrected to nearest value in 
list of preset values.
Density units
g/cm
3
Density cutoff
0.2 g/cm
3
0.0 – 0.5 g/cm
3
D1
0.00000
D2
1.00000
K1
1000.00
K2
50,000.00
FD
0.00000
Temp Coefficient
4.44
Slug flow
Slug flow low limit
0.0 g/cm
3
0.0 – 10.0 g/cm
3
Slug flow high limit
5.0 g/cm
3
0.0 – 10.0 g/cm
3
Slug duration
0.0 sec
0.0 – 60.0 sec
Temperature
Temperature damping
4.8 sec
0.0 – 80.0 sec
User-entered value is 
corrected to nearest lower 
value in list of preset values.
Temperature units
Deg C
Temperature calibration factor 1.00000T0.0000
T-Series sensor
D3
0.00000
D4
0.00000
K3
0.00000
K4
0.00000
FTG
0.00000
FFQ
0.00000
DTG
0.00000
DFQ1
0.00000
DFQ2
0.00000
Special units
Base mass unit
g
Base mass time
sec
Mass flow conversion factor
1.00000
Base volume unit
L
Base volume time
sec
Volume flow conversion factor 1.00000
Base GSV time
min
Base GSV unit
SCF
GSV conversion factor
1.00000
Variable 
mapping
Primary variable
Mass flow
Secondary variable
Density
Tertiary variable
Mass flow
Quaternary variable
Volume flow
Table A-1
Transmitter default values and ranges continued
Type
Setting
Default
Range
Comments