Emerson 4500 Manuel D’Utilisation

Page de 70
Reference Manual 
00809-0100-4027, Rev AA
December 2005
4-5
Rosemount 4500
Choosing a Trim 
Procedure
To decide which trim procedure to use, you must first determine whether the 
analog-to-digital section or the digital-to-analog section of the transmitter 
electronics need calibration. Refer to Figure 4-1 and perform the following 
procedure:
1. Connect a pressure source, a HART Communicator or AMS, and a 
digital readout device to the transmitter.
2. Establish communication between the transmitter and the HART 
Communicator.
3. Apply pressure equal to the upper range point pressure.
4. Compare the applied pressure to the Process Variable (PV) line on the 
HART Communicator On-line Menu or the Primary Variables screen in 
AMS. 
a. If the PV reading does not match the applied pressure (with 
high-accuracy test equipment), perform a sensor trim.
5. Compare the Analog Output (AO) line, on the HART Communicator or 
AMS, to the digital readout device. 
a. If the AO reading does not match the digital readout device (with 
high-accuracy test equipment), perform an output trim.
Sensor Trim
Trim the sensor using either full or zero trim functions. Trim functions vary in 
complexity and are application-dependent. Both trim functions alter the 
transmitter’s interpretation of the input signal.
Zero trim is a single-point adjustment. It is useful for compensating for 
mounting position effects and is most effective when performed with the 
transmitter installed in its final mounting position. Since this correction 
maintains the slope of the characterization curve, it should not be used in 
place of a full trim over the full sensor range.
When performing a zero trim, ensure that the equalizing valve is open and all 
wet legs are filled to the correct levels.
NOTE
Do not perform a zero trim on 4500 Absolute pressure transmitters. Zero trim 
is zero based, and absolute pressure transmitters reference absolute zero. To 
correct mounting position effects on a 4500 Absolute Pressure Transmitter, 
perform a low trim within the full sensor trim function. The low trim function 
provides a “zero” correction similar to the zero trim function, but it does not 
require zero-based input.
Full trim is a two-point sensor calibration where two end-point pressures are 
applied, and all output is linearized between them. Always adjust the low trim 
value first to establish the correct offset. Adjustment of the high trim value 
provides a slope correction to the characterization curve based on the low 
trim value. The factory-established characterization curve is not changed by 
this procedure. The trim values allow you to optimize performance over your 
specified measuring range at the calibration temperature.