 Model 2400S Transmitters for PROFIBUS-DP
Measurement Performance
10.3.3
Running Smart Meter Verification
To run a Smart Meter Verification test:
Using the display, see Figures 10-6 and 10-7.
Using a PROFIBUS host with the EDD, refer to Figure C-7 and follow the procedure 
illustrated in Figure 10-8.
Using PROFIBUS bus parameters, use the Diagnostic block (see Table D-4) and follow the 
procedure illustrated in Figure 10-9.
Note: If you start a Smart Meter Verification test using ProLink II, the EDD, or PROFIBUS bus 
parameters, and the outputs are set to Last Measured Value or Fault, the transmitter display shows 
the following message:
SENSOR
 
VERFY/x%
Table 10-3
PROFIBUS bus parameters interface for meter verification 
Step number
Step description
Interface
(1)
(1) For detailed information, see Table D-4
.
1
Set output state
Diagnostic block (Slot 3)
Index 54
2
Set uncertainty limit
Diagnostic block (Slot 3)
Index 55
3
Start/abort procedure
Diagnostic block (Slot 3)
Index 53
4
Check current algorithm state
Diagnostic block (Slot 3)
Index 56
5
Read percent complete
Diagnostic block (Slot 3)
Index 61
6
Check algorithm abort state
Diagnostic block (Slot 3)
Index 58
7
Check inlet stiffness
Diagnostic block (Slot 3)
Index 59
8
Check outlet stiffness
Diagnostic block (Slot 3)
Index 60
9
Read abort code
Diagnostic block (Slot 3)
Index 57