Emerson 51-DO-03 Manuel D’Utilisation
![Emerson](https://files.manualsbrain.com/attachments/1ef3aaad462e024fde6cecf537a4c6c781334d44/common/fit/150/50/6dab323528eefa0342baa74a10c15551c3b956250cf9d62f2117dcdf42a7/brand_logo.jpeg)
52
MODEL DO-03/04
SECTION 9.0
MAINTENANCE
9.3 AIR COMPRESSOR (PN 9380088 115Vac or PN 9380089 220 Vac)
The air compressor has no user-serviceable parts. It does not require lubrication. Fuse sizes are shown in the
table. The fuse is located below the T-shaped plastic cover next to the air compressor.
table. The fuse is located below the T-shaped plastic cover next to the air compressor.
33523-00
Electrolyte Fill Plug
9550094
O-Ring, Viton 2-014
24057-00
Membrane Retainer (smooth cap)
23501-00
Dissolved Oxygen Membrane Assembly: includes one membrane assembly and one O-ring
23502-00
Dissolved Oxygen Membrane Kit: includes 3 membrane assemblies and 3 O-rings
9210264
#1 Dissolved Oxygen Sensor Fill Solution, 4 oz (120 mL)
SPARE PARTS
FIGURE 9-1. Sensor Parts
Power
Fuse
100-115 Vac
AGC-4, 4A, 250V
230 Vac
AGC-2, 2A, 250V
DANGER
Hazardous Voltage
Can cause severe injury or death.
Disconnect power before servicing.
Disconnect power before servicing.