Emerson 51-DO-04 Manuel D’Utilisation

Page de 70
Output setpoints
Simulated tests
Configure
Exit
Enter
MODEL DO-03/04
SECTION 5.0
SOFTWARE CONFIGURATION
5.4 CHOOSING DISPLAY OPTIONS
4.
Move the cursor to "Configure" and press Enter (F4).
5. With the cursor on "Display", press Enter (F4). 
6.
The screen at left appears. The default settings shown for “Measure”
(Oxygen) and “Sensor” (RMT Standard) are appropriate for the
499ADO sensor. There are no changes to make.
If you wish to change the units, use the 
 key to move the cursor to
"Meas units" and press Edit (F4). Use the 
 or   keys to scroll
through the choices. Press Save (F4) to store the selection.
Display
Outputs
Alarms
Exit
Enter
Measure 
:  Oxygen
Sensor: RMT Standard
Meas units: ppm
Exit
Edit
1.  The 54eA analyzer can be used with other oxygen sensors manufactured by Rosemount Analytical as well as with
chlorine, monochloramine, and ozone sensors. The user must configure the analyzer to match the sensor being
used. 
2.  The display menu also lets the user customize the third line in the display, change timeout values, choose a lan-
guage other than English, and change the display contrast. 
3.
Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).
27