Emerson 51-DO-04 Manuel D’Utilisation

Page de 70
28
MODEL DO-03/04
SECTION 5.0
SOFTWARE CONFIGURATION
5.4 CHOOSING DISPLAY OPTIONS (CONTINUED)
7.  Set the remainder of the display parameters. Use the 
 and  keys
to choose the desired parameter. Then press Edit (F4). Use the 
key to move the cursor to the desired selection. Press Save (F4) to
store. 
SECURITY CAUTION
The analyzer uses the timeout value to activate security. Once
the analyzer is unlocked by entering a security code, security will
not re-activate until a display timeout occurs. If timeout has been
turned off, security will never reactivate.
Temp units: °C
Output 1:  mA
Output 2:  mA
Exit
Edit
Language:  English
Display left: I
Display right:  Out 2
Exit
Edit
Display contrast:  40
Timeout:  On
Timeout value:  10 min
Exit
Edit
Temp units
°C or °F
Output 1
mA or % of full scale
Output 2
mA or % of full scale
Language
English, Français, Español, Deutsch, Italiano
Display left
sensor current (I), alarm 1 setpoint (no units), 
alarm 3 setpoint (no units), or blank
Display right
sensor current (I), alarm 2 setpoint (no units), 
alarm 3 setpoint (no units), output 2, or blank
Display Contrast  00 (lightest)-99 (darkest); the display contrast 
changes as the number changes
Timeout
Timeout returns the display from any other screen 
to the main display if no key is pressed before the 
timeout value is exceeded.
Polling address
Identifies analyzer in multi-drop HART applications.
Timeout value