ResMed Stellar 150 Manuel D’Utilisation

Page de 40
13
Starting therapy
En
glis
h
Turning off the power
1
Stop the therapy.
2
Press the power switch 
 at the back of the device once and follow the instructions on 
the display.
Note: To disconnect the device from the mains power pull out the power cord from the 
power socket.
Working with alarms
WARNING
This device is not intended to be used for vital signs monitoring. If vital signs 
monitoring is required, a dedicated device should be used for this purpose.
This device is fitted with alarms to alert you to changes that will affect your treatment. 
Alarm messages are displayed along the top of the screen. High priority alarms are 
displayed in red, medium priority alarms in yellow and low priority alarms in light blue. The 
Alarm LED lights red during high priority alarms and yellow during medium and low priority 
alarms. 
The alarm volume can be set Low, Medium or High. From the Setup menu, select Alarm 
Settings
. After the set value has been confirmed, the alarm will sound and the alarm LED 
lights.
You can mute an alarm by pressing 
 once. By pressing the Alarm mute key again, the 
alarm sounds again. When an alarm is muted, the Alarm mute key LED will light constantly. 
For a high or medium priority alarm, if after two minutes the problem is still present, the 
alarm will sound again. Any active low priority alarm will be permanently muted and the 
Internal Battery Use alarm will be cleared until the alarm condition is fulfilled again.
Alarm message
Alarm LED
Alarm mute key