14
Tailoring treatment setup options
Setting Ramp
Ramp time is a feature that can be enabled 
by your clinician by setting a maximum ramp 
time. Designed to make the beginning of 
treatment more comfortable, ramp time is 
the period during which the pressure 
increases from a low start pressure to the 
treatment pressure. See “Setup menu: 
Options” on page 16.
Programs
If the Clinician has selected both, ie, dual, 
you can choose the program to use on the 
Treatment screen. If only a single program is 
selected, the option does not display.
Using mask-fit
You can use mask-fit to help you fit your mask properly. This feature delivers constant 
treatment pressure for a three-minute period, prior to starting treatment, during which you 
can check and adjust your mask-fit to minimise leaks. The mask-fit pressure is the set 
CPAP or EPAP pressure or 10 cm H
2
O, whichever is greater.
1
Put the mask on according to the mask user guide.
2
Hold down 
 for at least three seconds until pressure delivery starts. 
3
If necessary adjust the mask, mask cushion and headgear until you have a good mask fit. 
After three minutes treatment will begin. Mask-fit can be stopped at any time by pressing 
.
Notes: 
• Press   for at least three seconds during mask-fit to start the treatment immediately. 
• The mask-fitting feature is disabled when the mask type Trach is selected.
Ramp time
Program