MAINTENANCE AND TROUBLESHOOTING
28
Troubleshooting
NOTE: If problems are not remedied by the suggested means, please contact ICCI.
Lift
Scale
SYMPTOMS
PROBABLE CAUSE
SOLUTIONS
Seat and/or upper column feels loose.
Nylon bushing is loose.
Casters/brakes noisy or stiff.
Fluff or debris in bearings.
Clean the or replace the casters.
Electric actuator fails to lift when 
button is pressed.
Hand-control or actuator 
connector loose.
Battery is low.
Battery not connected properly 
to control box.
The connecting terminals are 
damaged.
Electric actuator in need of 
service or load is too high.
Check connections.
Reconnect the battery to the 
control box. Refer to Charging the 
Battery on page 16
.
Contact ICCI.
Unusual noise from actuator.
Actuator is worn or damaged or 
spindle is bent.
Contact ICCI.
SYMPTOMS
PROBABLE CAUSE
SOLUTIONS
Scale does NOT work properly.
Battery failure.
Check battery. Replace if necessary.
Battery has been replaced and 
scale still does NOT work properly.
Contact ICCI.