SECTION 6—TROUBLESHOOTING
20
SECTION 6—TROUBLESHOOTING
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.
SYMPTOMS
FAULTS
SOLUTION
Patient Lift feels loose.
Mast/Base joint loose.
Tie - Rods are loose.
Noisy or dry sound from pivots.
Needs lubrication.
Electric actuator fails to lift when button 
is pressed.
Hand-control or actuator connector 
loose.
Battery low.
RED emergency stop button pressed IN.
Battery not connected properly to con-
trol box.
The connecting terminals are damaged.
Electric actuator in need of service or 
load is too high.
Rotate RED emergency stop button 
CLOCKWISE until it pops out. 
Reconnect the battery to the control box. 
Refer to Charging the Battery on page 13.
Unusual noise from actuator.
Actuator is worn or damaged or spindle 
is bent.
Lift arms will not lower in uppermost 
position.
Lift arms require a minimum weight load 
to lower from the uppermost position.
Pull down slightly on the lift arms.
Lift arms will not lower during a power 
retraction.
Shoulder bolt at the junction of the boom 
and mast may not be properly installed.