Delta Faucet DF-745 Manuel D’Utilisation

Page de 98
46                                                                           DF-745
User Interface
Use the arrow keys (
and 
 to navigate the keyboard and use the Enter key to accept 
each character. If an error is made use the Next key to move to the Clear Entry key and hit 
enter. When the location name is complete use the Next key to move the highlight to 
Accept New Location and hit Enter. The display will return to the Downloader Screen 
and the name will appear in the list of available locations.
7.4.5.4 Delete Location
The delete location function is used to remove a location from the list of available names. 
On the Moisture Data Downloader screen Figure 35, use the Next key to move to the list of 
existing names and then use the arrow keys (
and 
 to select the location desired. 
Then use the Next key to move to Delete Location and press Enter. A confirmation box 
will then appear (see Figure 38) and the user can either accept the deleted selection with the 
Enter key or can hit ESC to cancel the action. If accepted, the name will be removed from 
the list of available locations.
Figure 38: Delete Selection
7.4.6 View Logs Menu
View Logs allows the user to easily access past events that may be connected with past 
operational changes (e.g., zero) or instrument upsets. Use the arrow keys (
and 
 to 
scroll up and down through the list. Pressing ESC will return to the main display.
7.4.6.1 Zero Log
Figure 39: View Zero Log Menu
The Moisture Zero Log reports on adjustments made to the moisture cell zero setting.  The 
date and time of the zero calibration is noted.  The zero is listed as either Manual or 
Automatic.  Notes are also given as to if the zero was aborted, failed due to timeout, or in