User Interface
Contact Delta F for assistance in interpreting the various codes if one should appear on the 
screen.
Figure 42: System Error Code Screen
7.4.6.3 Pump Capacity Test Log
Figure 43: Pump Capacity Test Log Menu
The pump capacity test is used to determine the condition of the aspirator or vacuum pump 
which in turn will have a direct impact the flow of gas through the analyzer. A reduction in 
pump capacity can result in a reduction in gas flow and as a result on the stability of the 
analyzer reading. See page Error! Bookmark not defined. for additional information on 
the pump capacity test. As the last step of the test, the system automatically puts an entry in 
the pump capacity test log for future reference. See Figure 43. Review of this information 
can be useful in detecting a trend in the condition of the aspirator or pump which can result 
in a need to rebuild the pump.