6888Xi Advanced Electronics Instruction Manual
Section 3: Configuration, Startup and Operation  
PN 51-6888Xi
December 2012
TX 
IOB  Parameter Name
Unit
Description
Y
Y
Maintenance 1
--
Alarm state: On/Off
O2 Sensor Open, O2 T/C Open, O2 Temp Low, O2 Temp High, 
O2 T/C Shorted, O2 T/C Reversed, Heater Failure, Burner Flameout
Y
Y
Maintenance 2
--
Alarm state: On/Off
Xmtr Disconnect, Cal Recommended, Cal Failed, Cell Imp High, 
Probe Mismatch
Y
Y
Advisory
--
Alarm state:
Cal Changed, Htr Volt Low, Probe Changed, Low O2, Diffuser Warning
N
Y
Previous Alarms
--
Previous Alarms (See Section 4, Troubleshooting)
N
Y
Ack Alarm Time
Days ago Time stamp of the previous alarms last acknowledged.
Y
Y
Duty Cycle
--
O2 heater duty cycle. Value between 0 and 1.
Y
Y
O2 Temp SP
degC
PID temperature set point
Y
Y
Heater Ramp Rate
degC/s
Heater ramp rate calculated in degree C per second.
Y
Y
Operating Status
--
Device configuration: On/Off
Factory Mode, Flame Status (IO Board DIP Switch setting), 
Auto Cal Device, Relay 1 Device, Relay 2 Device, AO Device
N
Y
I/O States 1
--
I/O state: On/Off
Gas 1 Solenoid, Gas 2 Solenoid
N
Y
I/O States 2
--
I/O state: On/Off
Flame Status In, Factory Mode, Relay1 Out, Relay2 Out, 
SPS/IMPS In, SPS/IMPS Out
Y
Y
Xmtr EE Val
--
Transmitter board nonvolatile memory diagnostic.
N
Y
IOB EE Val
--
I/O board nonvolatile memory diagnostic.
Y
Y
Xmtr Restart
--
Software restarts count for the Transmitter.
N
Y
IOB Restart
--
Software restarts count for the IO board.
Y
Y
HART | Device Rev
--
HART Device Revision number.
Y
Y
Transmitter | Version
--
Software version number for the Transmitter.
N
Y
I/O Board | Version
--
Software version number for the I/O board
Y
Y
Tag
--
Device tag: Up to 8 characters long.
Y
Y
Serial Number
--
Probe serial number.
Y
Y
Device ID
--
Unique Device ID number. (HART)  
Y
Y
PV
--
Primary variable assignment: O2 (HART)
Y
Y
SV
--
Secondary variable assignment: O2 Temp (HART)
Y
Y
TV
--
Third variable assignment: Cell Imp (HART)
Y
Y
4V
--
Fourth variable assignment: O2 Cell (HART)
N
Y
Poll Address
--
I/O board polling address
Y
Y
Xmtr Address
--
Transmitter board polling address
N
Y
T90 Filter
sec
Analog output T90 time. It represents the time to take a step change in
oxygen to reach 90% of the final value at the filter output.
N
Y
Low O2 Alm SP
%
Low O2 alarm threshold.
N
Y
O2 Cell Ref
%
O2 sensor reference gas percentage. It allows using a sensor reference gas
other than air. (Parameter for Programmable Reference feature.)
N
Y
High Temp Alm SP
degC
High temperature alarm threshold. (Parameter for Extended Temperature
feature.)
N
Y
Heater SP
--
Heater set point. (Parameter for Extended Temperature feature.)
0 = Normal (736C), 1 = Low (550C)
Table 3-5. Probe Parameters (2 of 4)