Configuration, Startup and Operation
47
6888Xi Advanced Electronics Instruction Manual
Section 3: Configuration, Startup and Operation  
PN 51-6888Xi
December 2012
TX 
IOB  Parameter Name
Unit
Description
Y
Y
Slope
mV/Dec
Current calibration slope. This is the slope value that was calculated as a result of
the last successful calibration.
Y
Y
Constant
mV
Current calibration constant. This is the constant value that was calculated as a
result of the last successful calibration.
Y
Y
Impedance
ohm
Cell Impedance. This is the sensor resistance that was calculated as a result of
the last successful calibration.
Y
Y
Time
Days ago Time stamp of the last successful calibration.
Y
Y
Cal Logs | Slope
mV/Dec
Previous calibration slope. There are ten calibration results. 1 is the most recent
and 10 is the least recent calibration slope
Y
Y
Cal Logs | Constant
mV
ration constant. There are ten calibration results. 1 is the most recent and 10 is
the least recent calibration constant.  
Y
Y
Cal Logs |
Impedance
ohm
Previous Cell Impedance. This is the sensor resistance that was calculated as a
result of previous successful calibration. . There are ten calibration results. Index
1 is the most recent and Index 10 is the least recent sensor resistance measured.
Y
Y
Cal Logs | Time
Days ago
Time stamp of the previous successful calibration. There are ten calibration time
stamp. Index 1 is the most recent and Index 10 is the least recent time stamp.
N
Y
Bad Slope
mV/Dec Failed calibration slope.
N
Y
Bad Constant
mV
Failed calibration constant.
Y
Y
Cal Result
--
Calibration result:
0 = None, 1 = Success, 2 = Failed Constant, 3 = Failed Slope, 
4 = Failed Temperature, 5 = Gas 1 Tolerance Error, 6 = Gas 2 Tolerance Error, 
7-10 (future), 11 = AutoCal No Resp, 12 = AutoCal OutofSync, 
13 = AutoCal Abort, 14 = No Solenoid, 15 = WarmUp Abort, 16 Alarm Abort
N
Y
Delta Imp
--
Delta impedance since last calibration.
N
Y
Cal Step
--
This represents the step of the calibration cycle is in:
0 = Idle, 1 = Reserved, 2 = Apply Gas 1, 3 = Flow Gas 1, 4 = Read Gas 1, 
5 = Apply Gas 2, 6 = Flow Gas 2, 7 = Read Gas 2, 8 = Cal Abort, 9 = Stop Gas, 
10 = Purge
N
Y
Time Remain
sec
Time remaining in the present calibration cycle state.
N
Y
Process
mV
Cell mV reading just prior to start of calibration. (Diagnostic parameter for
Diffuser Warning feature.)
N
Y
Diag State
--
The state of the Diffuser Warning state machine: (Diagnostic parameter for
Diffuser Warning feature.)
0 = Idle, 1 = Wait Gas 1, 2 = Stable Gas 1, 3 = Read Gas 1, 4 = Wait Gas 2, 
5 = Stable Gas 2, 6 = Read Gas 2, 7 = Wait Process, 8 = Rtn Process
N
Y
Init Resp
sec
The time it takes for the first calibration test gas to reach the sensor cell after the
solenoid is energized. (Diagnostic parameter for Diffuser Warning feature.)
N
Y
Process Resp
sec
The time it takes to see the initial response of the process after the gas 2 sole-
noid is turned off. (Diagnostic parameter for Diffuser Warning feature.)
N
Y
Rtn Process
sec
The time it takes from the initial response of the process until the process value
has stabilized close to the value of the process when the calibration started.
(Diagnostic parameter for Diffuser Warning feature.)
N
Y
O2 Rate Change
mV/s
The O2 cell mV rate of change. (Diagnostic parameter for Diffuser Warning 
feature.)
N
Y
Diff T90
sec
The time it takes from the gas-2 solenoid is turned off to return to 90% of the
process when the calibration started. (Diagnostic parameter for Diffuser
Warning feature.)
N
Y
Purge Time %
%
The percent of purge time used. This parameter can be used to estimate amount
of diffuser pluggage. (Diagnostic parameter for Diffuser Warning feature.)
N
Y
Diff Warn
--
The reason of the Diffuser Warning alarm: (Diagnostic parameter for Diffuser
Warning feature.)
0 = None, 1 = Gas 1 No Detect, 2 = Gas 2 No Detect, 3 = Gas 1 Unstable, 
4 = Gas 2 Unstable, 5 = Process No Detect, 6 = Purge End Premature, 
7 = Fail Reach Process, 8 = Diffuser Plugging, 9 = Constant Shifting
Table 3-5. Probe Parameters (4 of 4)