Configuring Custody Transfer
Figure 14-1
Security menu for custody transfer (NTEP)
14.5
Configuring custody transfer (OIML) and custody transfer (OIML/batch)
To configure custody transfer (OIML) or custody transfer (OIML/batch):
1. Referring to Figure 14-2:
a.
Set 
World Area
 to 
OIML
.
b.
Set 
Discrete Batch
 (if displayed) as desired:
Approved
: The discrete batch application can be used for transaction measurement 
and management. 
Not Approved
: The discrete batch application cannot be used for transaction 
measurement and management. 
c.
Set 
API
 or 
Enhanced Density
 (if displayed) as desired:
Approved
: API or enhanced density process variables can be used as the transfer 
variable.
Not Approved
: API or enhanced density process variables cannot be used as the 
transfer variable.
2. Ensure that the Series 3000 system, including all special applications (discrete batch, 
petroleum measurement, enhanced density), is completely configured and tested. Verify the 
following: 
Volume Flow Type
 is set to 
Liquid
. Gas standard volume flow cannot be used for OIML 
custody transfer.
If you will use the discrete batch application for transaction measurement, verify that the 
discrete batch flow source has been set to the process variable to be used for custody 
transfer measurement (the transfer variable). Only the following process variables are 
valid:
-
Mass flow
-
Volume flow (liquid)
-
API corrected volume flow (if the petroleum measurement application was approved 
in the previous step)
-
ED net mass flow (if the enhanced density application was approved in the previous 
step)
3. Perform a sensor zero. See Section 16.3.
World area
• NTEP
• OIML
Switch status
(1)
Security
Weights & Measures
(1) Read-only.