136
Micro Motion
®
 Series 3000 MVD Transmitters and Controllers
Startup Procedures
5. When the calibration is complete:
In the active alarm log, the alarm changes from “active” to “inactive but unacknowledged.” 
If the calibration succeeded, a “Success” status is displayed.
If the calibration failed, a “Failed” status is displayed.
6. Press 
EXIT
 to return to the Sensor Zero screen.
7. (Optional) To clear the alarm from the active alarm log and the alarm bar at the top of the 
display, acknowledge the alarm using the procedures described in Chapter 22.