Emerson Model 1700 Manuel D’Utilisation

Page de 350
Tip
When you reset a single totalizer, the values of other totalizers are not reset. Inventory values are not
reset.
8.7.1
Reset totalizers using the display (standard option)
Prerequisites
The Totalizer Reset display function must be enabled.
The totalizer that you want to reset must be configured as a display variable. For example:
If you want to reset the mass totalizer, Mass Total must be configured as a display
variable.
If you want to reset the volume totalizer, Volume Total must be configured as a display
variable.
Procedure
To reset the mass totalizer:
1. Scroll until the mass totalizer value appears.
2. Select.
3. Scroll until RESET appears beneath the current totalizer value.
4. Select.
5. Select again to confirm.
6. Scroll to EXIT.
7. Select.
To reset the volume totalizer:
1. Scroll until the volume totalizer value appears.
2. Select.
3. Scroll until RESET appears beneath the current totalizer value.
4. Select.
5. Select again to confirm.
6. Scroll to EXIT.
7. Select.
To reset the gas standard volume totalizer:
1. Scroll until the gas standard volume totalizer value appears.
2. Select.
3. Scroll until RESET appears beneath the current totalizer value.
4. Select.
5. Select again to confirm.
Transmitter operation
Configuration and Use Manual 
131