9.5.3 Procedure
1. Place  the  transmitter  in  ORP  mode.  See  Section  10.5.1,  steps  1  -  3. After
selecting and saving OrP, press  EXIT  twice to return to the main display.
2. Enter  the  CALIBRATE  menu  by  pressing    CAL    on  the  IRC.  The  Std sub-
menu appears.
3. Rinse the sensor with deionized water and place it in the ORP standard along
with a thermometer. Submerge the sensor tip at least three inches below the
surface  of  the  liquid.  Swirl  the  sensor  to  dislodge  trapped  air  bubbles. The
main display will show the measured ORP based on the previous calibration.
4. Once  the  temperature  and  ORP  readings  are  stable,  press  ENTER  .  The
screen changes to look like the figure to the left.
5. use the editing keys to change the flashing display to the desired ORP read-
ing. Press  ENTER  to save.
6. Press  EXIT  to return to the main display.
NOTE
A  transmitter  adjacent  to  the  one  being  calibrated  may  pick  up
signals  from  the  IRC.  To  avoid  accidentally  changing  settings,
use  a  different  security  code  for  each  nearby  transmitter.  See
Section 5.8, Security.
NOTE
during calibration, the sensor may be exposed to solutions hav-
ing  ORP  outside  the  normal  range  of  the  process.  To  prevent
false alarms and possible undesired operation of chemical dos-
ing  pumps,  place  the  analyzer  in  hold  during  calibration.  See
Section 9.3, using the Hold Function, for details.
CALIBRATE
Std
EXIT
NEXT       ENTER
CALIBRATE
Std
EXIT
NEXT       ENTER
0
2 5
.0
Std
9.5  Standardization
9.5.1 Purpose
This section describes how to prepare ORP standard solutions and how to make the transmitter reading match the ORP
of the standard. Procedures for making ORP standards are taken from ASTM Method d1498-93.
9.5.2 Preparation of ORP Standard Solutions
ASTM d 1498-93 gives procedures for making iron (II) - iron (III) and quinhydrone ORP standards. The iron (II) - iron (III)
standard is recommended. It is fairly easy to make and has a shelf life of about one year. In contrast, quinhydrone stan-
dards contain toxic quinhydrone and have only an 8-hour shelf life.
Iron (II) - iron (III) standard is available from Rosemount Analytical as PN R508-16OZ. The ORP of the standard solution
measured against a silver-silver chloride reference electrode is 476±20 mV at 25°C.
MODEL 5081-P PH/ORP
SECTION 9.0
CALIBRATION OF ORP MEASUREMENTS
1400
mV