MODEL 5081-P pH/ORP
SECTION 10.0
PROGRAMMING  FOR ORP MEASUREMENTS
ITEM
MNEMONIC
DISPLAy
FACTORy
USER 
LIMITS
SETTINGS
SETTINGS
PROGRAM LEvEL
A. Output Range (Section 10.3)
OutPut
(for Model 5081-P-HT only)
1. 4 mA Output 
-1400 to 1400 mV
-1400 mV
_______
2. 20 mA Output:
-1400 to 1400 mV
1400 mV
_______
3. Hold
HoLd
3.80 to 22.00 mA
21.00mA
_______
4. dampening
dPn
0 to 255 seconds
0 seconds
_______
5. Fault Current Output Setting
FAuLt
3.80 to 22.00 mA
22.00mA
_______
B. Diagnostic (Section 10.4)
dIAGnOStIC
1. Reference Cell Offset (Standardize error)
rOFFSt
0 to 1000 mV
60 mV
_______
2. diagnostics Function
dIAG
On/Off
Off
_______
3. Glass Impedance Temperature Correction)
IMPtC
On/Off
Off
_______
4. Reference Cell Impedance Type
rEF
LO/HI
LO
_______
5. Reference Cell High Impedance Fault 
rFH
0 to 2000 megohms (HI)
1500 megohms
_______
0 to 2000 kilohms     (LO)
140 kilohms
6. Reference Cell High Impedance Warning
rWH
0 to 2000 megohms (HI)
1000 megohms
_______
0 to 2000 kilohms     (LO)
40 kilohms
7. Reference Cell Low Impedance Warning
rWL
0 to 900 megohms   (HI)
20 megohms
_______
does not apply for low impedance reference cell
8. Reference Cell Low Impedance Fault 
rFL
0 to 900 megohms  (HI)
10 megohms
_______
does not apply for low impedance reference cell
C. Display (Section 10.6)
dISPLAY
1. Measurement type
tYPE
pH/ORP
pH
_______
2. Temperature units
tEMP
°C/°F
°C
_______
3. Output units (for Model 5081-P-HT only)
OutPut
mA/% of full scale
mA
_______
4. Code
COdE
0 to 999
000
_______
D. Output Simulation (Section 10.7)
SIMOutPut
(for Model 5081-P-HT only)
1. Test
tESt
3.80 to 22.00 mA
12.00 mA
_______
TABLE 10-1. ORP Settings LIst