7.1  GENERAL
The Calibrate menu allows the user to calibrate the pH and temperature response of the sensor. 
The transmitter does a two-point pH calibration. Both manual and auto calibration are available. In auto calibration the
transmitter automatically stores temperature-corrected calibration data once readings have met programmed stability lim-
its. In manual calibration the user enters buffer values and judges when readings are stable. The transmitter reading can
also be made to match the reading of a second pH meter. 
Temperature calibration is a one-point standardization against a reference thermometer. 
Prompts guide the user through the calibration procedures.
7.2.  ENTERING AND LEAvING THE CALIBRATE MENU
Press  CAL  on the infrared remote controller (IRC) to enter the Calibrate menu. To store new settings in memory, press
ENTER . To leave the Calibrate menu without storing new values, press  EXIT .  Pressing  EXIT  with a prompt showing
returns the display to the first prompt in the sub-menu. Pressing  EXIT  a second time returns the transmitter to the process
display.
If program settings are protected with a security code, pressing  PROG  or  CAL  will cause the Id screen to appear. Key
in the security code and press  ENTER . The first sub-menu will appear. For more information, see Section 5.8, Security.
A transmitter adjacent to the one being calibrated may pick up signals from the IRC. To avoid accidentally changing set-
tings, use a different security code for each nearby transmitter. See Section 5.8, Security, and Section 8.5, display units,
for details.
7.3   USING THE HOLD FUNCTION
during calibration, the sensor may be exposed to solutions having pH outside the normal range of the process. To prevent
false alarms and possible undesired operation of chemical dosing pumps, place the transmitter in hold during calibration.
Activating hold keeps the transmitter output at the last value or sends the output to a previously determined value.  
After calibration, reinstall the sensor in the process stream. Wait until readings have stabilized before deactivating Hold.
To activate or deactivate Hold, do the following:
1. Press  HOLd  on the IRC.
2. The HoLd prompt appears in the display. Press é or ê to toggle the Hold function between On and OFF.
3. Press  ENTER  to save.