Emerson 5081-P Manuel D’Utilisation

Page de 156
71
7.5  AUTO CALIBRATION
7.5.1 Purpose
1.  New sensors must be calibrated before use. Regular recalibration is also necessary.
2.  The use of auto calibration instead of manual calibration is strongly recommended. Auto calibration avoids com-
mon pitfalls and reduces errors.
3.  For more information about calibration in pH measurements and the use of buffers, refer to Section 8.7, Buffers and
Calibration.
7.5.2  what Happens During Auto Calibration?
1. The transmitter displays prompts that guide the user through a two-point buffer calibration.
2. The transmitter recognizes the buffers and uses the temperature-corrected pH value in the calibration. The trans-
mitter also measures noise and drift and does not accept calibration data until readings are stable. Stability limits
are user-programmable. See Section 8.7, Buffer Calibration Parameters.
7.5.3 Use of Calibration Standards (buffers)
1. A pH measurement is only as good as the calibration, and the calibration is only as good as the buffers used. A care-
ful buffer calibration is the first step in making an accurate pH measurement.
2. Calibrate with buffers having pH values that bracket the pH of the process. For example, if the pH is between 8 and
9, calibrate with pH 7 and 10 buffers. Commercial buffers for intermediate range pH are readily available. Buffers
outside the range pH 3.0 to pH 10.0 may not be readily available and must be prepared by the user. Tables 7-2 and
7-3 in Section 7.6, Buffer Calibration Parameters, list the buffers that the transmitter recognizes.
3. Allow time for the sensor and buffers to reach the same temperature. If the sensor was just removed from a process
having a temperature more than 10°C different from the buffer, allow at least 20 minutes.
4. For best results, calibrate with buffers having the same temperature as the process. If the buffer and process tem-
perature differ by more than about 15°C an error as great as 0.1pH may result.
5. Be careful using buffers at high temperatures. Protect the solution from evaporation. Evaporation changes the con-
centration of the buffer and its pH. Be sure the pH of the buffer is defined at high temperatures. Finally, no matter
what  the  temperature  is,  allow  the  entire  measurement  cell,  sensor  and  solution,  to  reach  constant  temperature
before calibrating.
6. The pH of a buffer changes with temperature. Equations relating pH to temperature for common buffers have been
programmed into the Model 5081-P pH transmitter. during auto calibration, the transmitter calculates the correct
buffer value and uses it in the calibration.
7. Buffers have limited shelf lives. do not use a buffer if the expiration date has passed. Store buffers at controlled room
temperature.
8. do not return used buffer to the stock bottle. discard it.
9. Protect buffers from excessive exposure to air. Atmospheric carbon dioxide lowers the pH of alkaline buffers. Other
trace gases commonly found in industrial environments, for example, ammonia and hydrogen chloride, also affect
the pH of buffers. Molds, from airborne spores, grow readily in neutral and slightly acidic buffers. Mold growth can
substantially alter the pH of a buffer.
10. Rinse the sensor with deionized water before placing it in a buffer. Remove excess water from the sensor by gently
daubing it with a clean tissue. do not wipe the sensor. Wiping may generate a static charge, leading to noisy read-
ings. The static charge may take hours to dissipate. A few drops of deionized water carried with the sensor into the
buffer will not appreciably alter the pH.
MODEL 5081-P pH/ORP
SECTION 7.0
CALIBRATION OF pH MEASUREMENTS