Emerson 5081-P Manuel D’Utilisation

Page de 156
76
MODEL 5081-P pH/ORP
SECTION 7.0
CALIBRATION OF pH MEASUREMENTS
CALIBRATE
Std
EXIT
NEXT       ENTER
CALIBRATE
SLOPE
EXIT
NEXT       ENTER
CALIBRATE
Std
EXIT
NEXT       ENTER
08
.00
5
9
.00
7.7.3 Procedure
1. Enter  the  CALIBRATE menu  by  pressing  CAL  on  the  IRC. The  CALibrAtE sub-menu
appears (pictured above left).
2. At the CALibrAtE sub-menu, press NEXT. The Std sub-menu appears.
3. With the Std sub-menu displayed, press ENTER. The Std prompt appears.
4. Be sure that the process pH and temperature are stable or, at worst, slowly drifting. Take
a grab sample from the process stream or sample line at a point as close as possible to
the pH sensor. Note the transmitter reading (pH
trans
) at the time the sample was taken.
5. Measure the pH of the sample (pH
std
) using the second pH meter. For best results make
the measurement at the same temperature as the process.
6. Note the current process reading (pH
curr
). Calculate the corrected reading from the equa-
tion:
pH
corr
= pH
curr
+ (pH
std
- pH
trans
)
where, pH
corr
is the corrected pH value, pH
curr
is the current process reading, pH
std 
is
the pH measured using the standard instrument, and pH
trans
is the pH measured by the
transmitter when the sample was taken. use the editing keys to change the flashing dis-
play to pH
corr
calculated above. Press  ENTER  to save the corrected pH. 
7. The transmitter converts the difference between pHcorr and pHcurr into mV and com-
pares  the  result  with  the  value  programmed  for  rOFFSt in  Section  8.4,  diagnostic
Parameters.  If  the  difference  exceeds  the  value  for  rOFFSt,  the  transmitter  will  not
accept the data and will not update the display to the corrected pH. The message StD
Err 
will appear. 
8. If the corrected pH value is acceptable, the display will change to look like the screen at
the left. The slope displayed is the current electrode slope. If the slope is incorrect and
the correct value is known, use the editing keys to change the slope to the desired value.
Press  ENTER  to save the value. To leave the slope unchanged, press  EXIT .
9. To leave the CALIBRATE menu, press  EXIT . 
NOTE
A transmitter adjacent to the one being calibrated may pick up signals from the IRC.
To avoid accidentally changing settings, use a different security code for each near-
by transmitter. See Section 5.8, Security.