Emerson 5081-P Manuel D’Utilisation

Page de 156
77
MODEL 5081-P pH/ORP
SECTION 8.0
PROGRAMMING FOR pH MEASUREMENTS
SECTION 8.0
PROGRAMMING FOR pH MEASUREMENTS
8.1
General
8.2
Entering and Leaving the Program Menu
8.3
Output Ranging
8.4
Diagnostic Parameters
8.5
Temperature Related Settings
8.6
Display Units
8.7
Buffer Calibration Parameters
8.8
Isopotential Parameters
8.9
Generating a Test Current
8.1 GENERAL
This section describes how to do the following:
1. assign pH values to the 4 and 20 mA outputs (for Model 5081-P-HT only), 
2. set the current generated by the transmitter during hold,
3. set the current generated by the transmitter when a fault is detected (for Model 5081-P-HT only),
4. change sensor diagnostic limits,
5. enable and disable automatic temperature compensation,
6. change the units of the displayed variables,
7. program a security code,
8. identify buffers for auto calibration,
9. change the transmitter isopotential point,
10. simulate output currents for testing (for Model 5081-P-HT only).
Factory default settings are given in Table 8-1. If default settings are acceptable, the transmitter is ready for calibration.
See Section 7.0, Calibration of pH Measurements. Once a setting has been changed, there is no way to automatically
reset the transmitter to factory defaults. Settings must be returned to default values one at a time. Figure 5-4 shows the
menu tree.
8.2 ENTERING AND LEAvING THE PROGRAM MENU
Press    PROG    on  the  infrared  remote  controller  (IRC)  to  enter  the  Program  menu.  To  save  new  settings,  press
ENTER. To leave the Program menu without saving new values, press  EXIT . Pressing  EXIT  with a prompt showing
returns the display to the first prompt in the sub-menu. Pressing  EXIT  again returns the transmitter to the process
display.
If program settings are protected with a security code, pressing  PROG  or  CAL  will cause the Id screen to appear.
Key in the security code and press  ENTER . The first sub-menu will appear. For more information, see Section 5.7,
Security.
A transmitter adjacent to the one being programmed may pick up signals from the IRC.  To avoid accidentally changing
settings, use a different security code for each nearby transmitter.  See Section 5.8, Security, and Section 8.6, display
units, for details.