PROGRAMMING FOR pH MEASUREMENTS
8.5.3 Procedure
1. Press PROG on the infrared remote controller (IRC).
2. Press NEXT until the tEMP sub-menu appears in the display. Press ENTER .
3. The screen displays the tAUTO prompt. Press é or ê to enable (On) or disable
(OFF) automatic temperature compensation. Press ENTER to save.
4. The tMAN prompt appears. use the editing keys to change the temperature to the
desired value. To enter a negative number, press è or ç until no digit is flashing.
Then press é or ê to display the negative sign. Permitted values are between -5.0
Then press é or ê to display the negative sign. Permitted values are between -5.0
and 130.0°C. If tAUTO was disabled in step 3, the temperature entered in this step
will be used in all subsequent measurements, no matter what the process tempera-
ture is. Press ENTER to save.
will be used in all subsequent measurements, no matter what the process tempera-
ture is. Press ENTER to save.
PROGRAM
tAUtO
EXIT
ENTER
ON
PROGRAM
tMAn
EXIT
ENTER
0
25
.0
PROGRAM
tEMP
EXIT
NEXT ENTER