Emerson 5081-P Manuel D’Utilisation

Page de 156
85
MODEL 5081-P pH/ORP
SECTION 8.0
PROGRAMMING FOR pH MEASUREMENTS
8.5.3 Procedure
1. Press  PROG  on the infrared remote controller (IRC).
2. Press  NEXT  until the tEMP sub-menu appears in the display. Press  ENTER .
3. The screen displays the tAUTO prompt. Press é or ê to enable (On) or disable
(OFF) automatic temperature compensation. Press  ENTER  to save.
4. The tMAN prompt appears. use the editing keys to change the temperature to the
desired value. To enter a negative number, press è or  ç until no digit is flashing.
Then press é or ê to display the negative sign. Permitted values are between -5.0
and 130.0°C. If tAUTO was disabled in step 3, the temperature entered in this step
will be used in all subsequent measurements, no matter what the process tempera-
ture is. Press  ENTER  to save.
PROGRAM
tAUtO
EXIT
ENTER
ON
PROGRAM
tMAn
EXIT
ENTER
0
25
.0
PROGRAM
tEMP
EXIT
NEXT       ENTER