Emerson 00825-0100-4530 Rev EC Manuel D’Utilisation

Page de 36
March 2013
4
Quick Installation Guide
Step 1: Confirm system readiness
Confirm HART revision capability
If using HART-based control or asset management systems, confirm the HART 
capability of those systems prior to transmitter installation. Not all systems 
are capable of communicating with the HART Revision 7 protocol. 
This transmitter can be configured for either HART Revision 5 or 7. 
For instructions on how to change the HART revision of your transmitter, see 
.
Confirm correct Device Driver
Verify that the latest Device Driver (DD/DTM) is loaded on your systems to 
ensure proper communication.
Download the latest Device Driver from 
.
Rosemount 5300 device revisions and drivers
 provides the information necessary to ensure that you have the correct 
Device Driver and documentation for your device.
Table 1.  Rosemount 5300 device revisions and files
Switch HART revision mode
If the HART configuration tool is not capable of communicating with HART 
Revision 7, the Rosemount 5300 Series will load a Generic Menu with limited 
capability. The following procedures will switch the HART revision mode from the 
Generic Menu:
1. Manual Setup > Device Information > Identification > Message
To change to HART Revision 5, Enter: “HART5” in the Message field
To change to HART Revision 7, Enter: “HART7” in the Message field
Firmware 
Version
1
1.   Firmware version is printed on the transmitter head label, e.g. SW 2E0.
Find Device Driver
Review Instructions
Review Functionality
HART 
Universal 
Revision
Device 
Revision
2
2.   Device revision is printed on the transmitter head label, e.g. HART Dev Rev 4.
Manual Document Number
Changes to Software
3
3.   HART Revision 5 and 7 are selectable.
2F0 and later
7
4
00809-0100-4530 
Rev EA
See footnote 3 for list of 
changes.
5
3
2A2 - 2D2
5
3
00809-0100-4530 
Rev DA
N/A