Grounding the flowmeter components
Grounding requirements depend on the installation type (see Figure 2-1
Grounding methods for 
each flowmeter component are listed in Table 2-3.
If national standards are not in effect, follow these grounding guidelines:
Use copper wire, 14 AWG (2,5 mm
2
) or larger wire size, for grounding.
Keep all ground leads as short as possible, less than 1 
Ω impedance.
Connect ground leads directly to earth, or follow plant standards.
CAUTION
Improper grounding could cause measurement error.
To reduce the risk of measurement error:
Ground the transmitter to earth, or follow ground network requirements for the 
facility.
For installation in an area that requires intrinsic safety, refer to Micro Motion 
approval documentation, shipped with the transmitter or available from the Micro 
Motion web site.
For hazardous area installations in Europe, refer to standard EN 60079-14 if 
national standards do not apply.
Table 2-3
Grounding methods for flowmeter components
Installation 
architecture
Components
Grounding method
Integral
Sensor / core processor / 
transmitter
Ground via piping, if possible (see sensor documentation). Otherwise, 
ground according to applicable local standards using either the 
transmitter’s internal or external ground screw.
4-wire remote
Sensor / core processor 
assembly
See sensor documentation.
Transmitter
Ground according to applicable local standards, using either the 
transmitter’s internal or external ground screw. 
9-wire remote
Sensor / junction box
See sensor documentation.
Transmitter / core 
processor assembly
Ground according to applicable local standards, using either the 
transmitter’s internal or external ground screw, or the core processor’s 
internal ground screw.
Remote core 
processor with 
remote 
transmitter
Sensor
See sensor documentation.
Core processor
Ground according to applicable local standards, using either the internal 
or external ground screw. 
Transmitter
Ground according to applicable local standards, using either the 
transmitter’s internal or external ground screw.