Emerson IFT9701 Manuel D’Utilisation

Page de 90
58
Model IFT9701 Transmitter Instruction Manual
Troubleshooting
 continued
Figure 8-1
Transmitter terminals and diagnostic LED
8.3.2
Optional LCD
Except for ATEX Zone 1 areas, the transmitter is available with an optional LCD. The 5-digit LCD is 
on the cover of the field wiring compartment, and indicates the mass or volume flow rate or the 
operating condition of the flowmeter. Table 8-2 explains the operating conditions indicated by the 
LCD.
8.3.3
Fault outputs
The transmitter produces downscale or upscale outputs to indicate a fault.
Downscale: The milliamp output goes to 2 mA; the pulse output goes to 0 Hz.
Upscale: The milliamp output goes to 22 mA; the pulse output goes to 7200 Hz.
Unless otherwise specified on the order, jumpers are set so the transmitter generates downscale fault 
outputs. To change jumper settings for fault outputs, see Section 2.3.
Table 8-1
Normal conditions indicated by LED
Diagnostic LED does this:
Condition
Blinks ON once per second (25% ON, 75% OFF)
Normal operation
Remains ON
Startup and initialization, zero in progress, transmitter hardware 
failure
Blinks OFF once per second (75% ON, 25% OFF)
Slug flow (density below or above user-defined limits)
Blinks ON 4 times per second
Fault condition
Remains OFF
Power-supply failure, transmitter hardware failure
Power supply
• If transmitter fails to 
produce outputs, 
check for reversed 
polarity
• Make sure actual 
power supply 
matches power 
supply shown on 
label
• AC power applied to 
DC transmitter will 
damage transmitter
• Power-supply fuses 
are not replaceable
Ground screw
for power supply
Case ground
Diagnostic LED
Shields (GND) for wiring to sensor 
(remote-mount transmitters only)
Sensor wiring 
compartment
(intrinsically safe)
Pulse output 
terminals
Milliamp output
terminals
+
20–30V
or
N
L
85–250V~
Intrinsically safe 
terminal block