Teledyne 701 Manuel D’Utilisation

Page de 69
 
01671H (DCN6051) 
3-1 
 
3 GETTING STARTED 
3.1 Unpacking 
 
CAUTION 
To avoid personal injury, always use two persons to lift and 
carry the Model 701
 
 
1.  Verify that there is no shipping damage. If there are signs of damage, immediately 
advise the shipper, then Teledyne API. 
2.  Remove the M701 from its shipping carton. 
3.  Remove the cover and check for damage inside. 
4.  Remove the four red shipping screws holding the compressor base to the bottom of 
the chassis. These screws are inside the chassis at the bottom, underneath the 
pump. See Figure 5-3. 
5.  Check the line voltage and frequency label on the rear panel and verify that it 
matches your local power.  
3.2 Installation 
The basic M701 has rubber feet for counter-top use. Optionally, the M701 is supplied 
with slides and angles for mounting in a standard 19” RETMA rack.  
When installing the M701, check the following points: 
1.  At least 6” (150 mm) clearance at the back for proper fan operation. 
2.  At least 3” (75 mm) at each side for proper venting. 
3.  Adequate support for the weight; 55 lb. (25 kg.). 
4.  Access at the rear for making the pneumatic and electrical connections. 
5.  Access at the front for occasional scrubber replacement. 
6.  For safety, the M701 should be operated with the cover on.