Teledyne 701 Manuel D’Utilisation

Page de 69
 
01671H (DCN6051) 
2-1 
 
2 SPECIFICATIONS AND WARRANTY 
2.1 Specifications 
Table 2-1.  Specifications 
PARAMETER SPECIFICATION 
Output 
STD 20 SLPM @ 30 psig 
 
Max delivery pressure 
35 psig  
(Except in some specials that have restricted output) 
Dewpoint 
< 15 SLPM 
 15 SLPM 
 
< -20
C 
< -10
C
 
SO
<0.5 ppb 
NO <0.5 
ppb 
NO
2
 <0.5 
ppb 
O
3
 <0.5 
ppb 
HC/CO 
Options < 0.02 ppm HC / < 0.025 ppm CO 
Purge air 
As required by the calibrator 
Compressor Single-cylinder, 
internal 
Mounting 
Bench type (standard) 
Rack mount (optional) 
Dimensions 
8.75” H x 17”W x 23.5”D  
(22.2 cm x 43.2 cm x 60 cm) 
Weight 
55 lbs. (25 Kg) 
Operating Temp 
5-40
C 
Power 
115V, 60Hz, 3.5A 
230V, 50Hz, 2.0A 
220V, 60 Hz, 2.2A 
Environmental 
Conditions 
Installation Category 
(Over Voltage Category ) II 
Pollution Degree 2