   9 CHECK AND ADJUSTMENT
9.3.4 Collimation of the Instrument
Collimation is required to make the line of sight of the telescope perpendicular to the horizontal axis of 
the instrument, otherwise, it will not be possible to extend a straight line by direct means.
l
Check
1) Set the instrument up with clear 
sights of about 50 to 60meters
(160 to 200 ft.) on both sides of the 
instrument.
2) Level the instrument properly with the 
plate level.
3) Sight Point A at approximately 50 
meters (160 ft.) distance.
4) Rotate the telescope 180° or 200g 
around the horizontal axis, so that the 
telescope is pointed in the opposite 
direction.
5) Sight Point B, at equal distance as 
Point A.
6) Rotate the instrument 180° or 200g 
around the vertical axis. Fix a sight 
on Point A once more.
7) Rotate the telescope 180°
or 200g around the horizontal axis 
once more and fix a sight on Point C, 
which should coincide with previous 
Point B.
8) If Points B and C do not coincide, 
adjust in the following manner.
l
Adjustment
1) Unscrew the cross-hair adjustment section  cover.
2) Find Point D at a point between Points C and B, which should
be equal to 1/4th the distance between Points B and C and 
measured from Point C. This is because the apparent error 
between Points B and C is four times the actual error since
the telescope has been reversed twice during the checking 
operation.
3) Shift the vertical cross-hair line and coincide it with
Point D, by revolving the left and right capstan 
adjustment screws with the adjusting pin.Upon 
completing the adjustment, repeat the checking 
operation once more.
If Points B and C coincide, further adjustment is not 
required. Otherwise, repeat the adjustment.
Eyepiece
Telescope
Capstan
adjustment
screws
Eyepiece
Capstan
adjustment
screws