   9 CHECK AND ADJUSTMENT
9.4
Adjustment of Compensation Systematic Error of Instrument
1) Error of vertical axis (X,Y tilt sensor offset)         2) Collimation error  
3) Error of vertical angle 0 datum                            4) Error of horizontal axis 
The above mentioned errors will be compensated by software, which calculated internally according to 
each compensation value.
Also these errors can be compensated by software collimating one side of the telescope that is carried 
out to delete the error by turning in normal and reverse both sides of telescope so far.
Operating procedure
Operation
Display
1
Level the instrument properly with the plate 
level.
2
Press  [F5] key from the main menu.
[F5]
3
Press  [F1] key .
[F1]
4
Collimate target A (around 0
°
 in horizontal 
within ±3
°
) in normal telescope setting 
(FACE(1)).
Collimate
 A
(Normal)
5
Press [F6](SET)key. *1)
The sample display shows that the 
measurement is made 5 times in FACE 1.
[F6]
6
Turn the telescope in reverse telescope 
setting (FACE(2)).
Turn 
telescope
Adjustment                     
F1 V0/Axis (Measurement)
F2 V0/Axis (Constant list)
F3 Date Time
F4 Instrument constant  
   ↓
    
    [V0/Axis Adjustments]
ERROR CORRECTION
(A) Tilt,V0 init,Collimation
(B) H Axis
       
      (A)COLLIMATION
           FACE 1
   V: 88°40'20"
    
 SKIP                   SET
LEVEL±0
/0
           FACE 1
   V: 89°55'50"
    
 SKIP                   SET
LEVEL±0
/5
           FACE 2
   V:270°04'20"
    
 SKIP                   SET
LEVEL±0
0/5