Siemens LU01 Manuel D’Utilisation

Page de 114
Page 10
SITRANS LU01 – INSTRUCTION MANUAL
7ML19985BE02
mmmm
m
Inst
allat
ion
Mounting
 
Inspect all cartons and packaging for possible damage during shipment, before removing 
the SITRANS LU01 and associated equipment. 
1.
Loosen the 6 enclosure lid (captivated) screws and swing the lid open. 
2.
Remove the 4 Board B mounting screws (outer corners) and remove the circuit 
board assembly. 
3.
Drill required holes in the enclosure bottom to meet enclosure cable/conduit entry 
requirements. 
4.
Attach the enclosure to the selected mounting surface (use four predrilled screw 
holes).
5.
Attach the conduits/cable hubs to the enclosure. (Do not apply undue force.) 
6.
Reinstall the circuit board assembly. 
Transducer Mounting 
Objects near the transducer face cannot be reliably detected. Mount the transducer 
above the highest material level (away from the nearest monitored object) by the 
following 
Nearest Distance
Warnings
Non metallic enclosure does not provide grounding between connections. Use 
grounding type bushings and jumpers. 
This product is susceptable to electrostatic shock. Follow proper grounding 
procedures. 
Nearest Distance
Transducer Types 
0.5 m (1.65 ft)
1
1.
 
This is the recommended minimum distance. However, it can be reduced under certain circum-
stances. Please check the appropriate transducer manual for details.
ST-H, ST-25, XRS-5, XCT-8, XCT-12, XPS-10, XPS-15, ST-50
0.66 m (2.17 ft)
XPS-30, XPS-40 
0.99 m (3.25 ft)
ST-100, LR-21, XLT-30, XLS-30 
1.32m (4.33 ft)
LR-13, XLT-60, XLS-60
1
2
3
4
1 2
mA
5
9
C
P
P
X X X
6
0
7
8
1
2
3
4
267 mm
(10.5")
285 mm
(11.2")
lid screw
(6 places)
209 mm
(8.2")
172 mm 
(6.8")
suitable location for conduit entrances.
Use water tight conduit hubs to maintain the enclosure rating.
mounting hole, 4.3 mm (0.17")
diameter, access under lid (4 places).
programmer
lid (hinged)
customer
mounting screws
106 mm
(4.2")