National Instruments SCXI-1121 Manuel D’Utilisation

Page de 162
Chapter 2
Configuration and Installation
© National Instruments Corporation
2-25
SCXI-1321 Offset-Null and Shunt-Calibration Terminal Block
The SCXI-1321 terminal block operates only with Revision C and later 
SCXI-1121 modules.
In addition to the 18 screw terminals, the SCXI-1321 has circuitry for 
offset-null adjust of Wheatstone bridges as well as a shunt resistor for 
strain-gauge shunt calibration. This terminal block works especially well 
with bridge-type transducers such as strain gauges. The SCXI-1321 can 
also easily accommodate thermocouples, RTDs, thermistors, millivolt 
sources, volt sources, and current-loop receivers.
SCXI-1321 Nulling Circuitry
The nulling circuitry operates with full-bridge, half-bridge, quarter-bridge, 
and strain-gauge configurations. Each channel has its own nulling circuitry 
and its own trimming potentiometer as listed in Table 2-10.
To null the static offset voltage of the bridge, use the following procedure:
1.
Configure your bridge to the selected channel.
2.
Select and read the channel output.
3.
While monitoring the output, rotate the trimmer wiper with a flathead 
screwdriver until you reach 0 V.
You have nulled your bridge and are ready for a measurement.
The nulling range for your terminal block is ±2.5 mV, assuming that you 
have a 120 
Ω
 strain gauge and 3.333 V excitation voltage. You can change 
this range by replacing the nulling resistor with a resistor of another value. 
Each channel has an independent nulling resistor. You can therefore mix 
your ranges to accommodate each channel requirement. Table 2-11 lists the 
nulling resistors and their corresponding channels.
Table 2-10.  Trimmer Potentiometer and Corresponding Channel
Channel Number
Trimmer Potentiometer
0
R1
1
R2
2
R14
3
R15