National Instruments SCXI-1121 Manuel D’Utilisation

Page de 162
Chapter 2
Configuration and Installation
2-26
www.natinst.com
The value of all the nulling resistors on your terminal block is 39 k
Ω
. Notice 
that these resistors are socketed for easy replacement. These sockets best fit 
a 1/4 W resistor lead size.
To determine your nulling range, use the following formula (refer to 
Figure 2-9 for visual help):
where
R
g
 is the nominal strain-gauge resistance value.
R
d
 is either a completion resistor or a second strain-gauge nominal 
resistance.
R
null
 is the nulling resistor value.
V
exc
 is the excitation voltage (3.333 or 10 V).
For example, assuming:
V
exc
 = 3.333 V
R
g 
= 120 
Ω
R
d
 = 120 
Ω
R
null
 = 39 k
Ω
V
nulling
 = ±2.56 mV
Table 2-11.  Nulling Resistors and Corresponding Channel
Channel Number
Nulling Resistor
0
R3
1
R5
2
R7
3
R9
V
nullingrange
V
exc
2
----------
V
exc
R
d
R
null
R
g
+
(
)
R
null
R
g
R
d
R
null
R
g
+
(
)
+
-----------------------------------------------------------
±
=