National Instruments SCC-AI05 Manuel D’Utilisation

Page de 18
© National Instruments Corporation
13
SCC-AI Series Isolated Analog Input Modules
Environmental
Operating temperature............................ 0 to 50 
°C
Storage temperature ............................... –20 to 70
 °C
Humidity ................................................ 10 to 90% RH, noncondensing
Maximum altitude .................................. 2,000 meters
Pollution Degree (indoor use only) ........ 2
Safety
The SCC-AIXX
 
meets the requirements of the following standards 
for safety and electrical equipment for measurement, control, and 
laboratory use:
IEC 61010-1, EN 61010-1
UL 3111-1, UL 61010B-1
CAN/CSA C22.2 No. 1010.1
Note
For UL and other safety certifications, refer to the product label, or visit 
ni.com/hardref.nsf
, search by model number or product line, and click the 
appropriate link in the Certification column.
Electromagnetic Compatibility
Emissions ............................................... EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
Immunity ................................................ Evaluated to EN 61326:1997/ 
A2:2001, Table 1
CE, C-Tick and FCC Part 15 (Class A) Compliant
Note
For full EMC compliance, you must operate this device with shielded cabling. 
In addition, all covers and filler panels must be installed. 
CE Compliance
The SCC-AIXX meets the essential requirements of applicable European 
Directives, as amended for CE marking, as follows:
Note
Refer to the Declaration of Conformity (DoC) for this product for any additional 
regulatory compliance information. To obtain the DoC for this product, visit 
ni.com/hardref.nsf
, search by model number or product line, and click the 
appropriate link in the Certification column.