Calibrating Gain and Offset Errors
The SCC-AIXX is calibrated at the factory before shipment. If you want to 
adjust the gain of the SCC-AIXX in your system using your E Series DAQ 
device, you need a voltage source that is several times more accurate than 
the SCC itself and capable of providing the DC voltage shown in Table 1-3.
To adjust the gain of the SCC-AIXX, complete the following steps:
1.
Select the desired SCC-AIXX channel on the E Series DAQ device.
2.
Set the gain on the E Series DAQ device so that its input range is 
±10 V.
3.
Connect the voltage source to the screw terminals of the desired 
channel on the SCC-AIXX.
4.
Apply the voltage given in Table 1-3 that corresponds to the 
SCC-AIXX module. For example, if you have an SCC-AI03, you must 
apply 8 VDC.
5.
Using your software, have the E Series DAQ device read the desired 
channel on the SCC-AIXX and record the value.
6.
Input 0 VDC to the SCC-AIXX.
7.
Use the E Series DAQ device to read that channel and record the value.
8.
Subtract the values read (first reading – second reading).
Table 1-3.  SCC-AIXX Input Voltage Requirements
Module
Input Range
SCC-AI01
40 V
SCC-AI02
16 V
SCC-AI03
8 V
SCC-AI04
4 V
SCC-AI05
800 mV
SCC-AI06
80 mV
SCC-AI07
40 mV
SCC-AI13
8 V
SCC-AI14
4 V