Chapter 4
Initial Transmitter Startup Procedures
4.1
Overview
The procedures described in this chapter should be performed the first time a transmitter is started. 
You can use ProLink II, the HART Communicator, AMS software, or the display to perform the 
procedures: the communications method does not matter.
The following procedures are described:
Using ProLink II to perform a loop test on transmitter outputs (and inputs, if your transmitter 
has a discrete input or frequency input)
Using ProLink II to trim the mA outputs
Using ProLink II to zero the meter
Note: The procedures in this chapter provide general methods for using ProLink II with your 
transmitter. For information on using the HART Communicator or the display, or for information 
specific to your transmitter such as the number and type of outputs, specific ranges for each output, 
etc., refer to the appropriate transmitter manual. Transmitter manuals are shipped with the 
transmitter, and are also available on the Micro Motion web site.
Note: Sections 4.2 and 4.3 do not apply to Series 2000 transmitters with Profibus-PA or F
OUNDATION
 
fieldbus.
4.2
Loop tests
loop test is a means to:
Verify that outputs are being sent by the transmitter and received accurately by the receiving 
devices
Determine whether or not you need to trim the mA outputs
Verify that the discrete input or frequency input sent by an external device is being received 
correctly by the transmitter (if the transmitter has a discrete input or frequency input)
The ProLink II loop test options are shown in Figure 4-1. Different options are available with different 
transmitters.