66
ProLink
®
 II Software for Micro Motion
®
 Transmitters
Transmitter Configuration, Characterization, and Calibration
5.8
Configuring polling
Polling is used to retrieve temperature or pressure data from an external device. These data can then 
be used for API calculation or other process variable calculations in applications that require pressure 
or temperature compensation. You may query one or two external devices.
You must also ensure that the primary mA output has been wired for HART protocol. See Chapter 2 
or the installation manual for your transmitter.
To configure polling:
1. From the 
View
 menu, select 
Preferences
.
a.
To use the external pressure data, ensure that the 
Enable External Pressure 
Compensation
 check box is checked.
b.
To use the external temperature data, ensure that the 
Use External Temperature
 check 
box is checked.
c.
Click 
Apply
.
2. Ensure that the 
External Temperature
 box in the 
Temperature
 tab has not been set. The 
default value is 32.00 °F (0.0000 °C).
3. Click the 
Polled Variables 
tab in the 
Configuration
 panel.
4. For Polled Variable 1:
a.
Click the arrow in the 
Polling Control 
box, and select a HART polling method from the 
list. Choose 
Poll DP HART Primary
 if the external device might be accessed by another 
device acting as a secondary master (e.g., a HART Communicator). Choose 
Poll DP 
HART Secondary
 if the external device might be accessed by another device acting as a 
primary master.
b.
Click 
Apply
. Polling parameters are now displayed for configuration.
c.
In the 
External Tag
 box, type the HART tag of the external device that you will poll.
d. Click the arrow in the 
Variable Type 
box, and select a process variable.
e.
Click 
Apply
.
f.
The value displayed in the 
Current Value
 textbox is the value received from the external 
device. Verify the value.
5. Repeat these steps for Polled Variable 2, if required.
Note: For some transmitters, the Polling Control selection for Polled Variable 1 is automatically used 
for Polled Variable 2. No configuration is required.