5. If you will use a static pressure value:
a.
Type the pressure value in the 
External Pressure
 box.
b.
Click 
Apply
.
c.
Ensure that neither polled variable is configured to poll for pressure. Polling for 
temperature is allowed. See Section 5.8.
5.7
Compensating for temperature
Temperature data are used in several different calculations. Micro Motion sensors always report 
temperature data to the transmitter. For greater accuracy, you can configure the transmitter to use a 
different temperature value:
If the temperature is a known static value, you may choose to enter the external temperature in 
the software and not poll a temperature measurement device.
If the operating temperature varies significantly, you may choose to have the transmitter poll 
for an updated temperature value from an external temperature measurement device. Polling 
requires HART protocol.
Note: If you have core processor v2.1 or earlier, the external temperature data are used for all 
calculations that require temperature or values. If you have core processor v2.2 or later, the external 
temperature data are used only for calculation of the derived variable in enhanced density 
applications or the CTL (Correction for Temperature on volume of Liquids) value in petroleum 
measurement applications.
Note: If you poll for temperature, ensure that the external temperature measurement device is 
accurate and reliable, and ensure that the transmitter is configured to use the same temperature unit 
that the external temperature measurement device is using.
To enable and configure temperature compensation:
1. From the 
View
 menu, select 
Preferences
 and ensure that the 
Use External Temperature
 
checkbox is checked.
2. If you will poll an external device for temperature data, follow the polling setup instructions in 
3. If you will use a static temperature value:
a.
Open the 
Configuration
 panel and click the 
Temperature 
tab.
b.
Type the temperature value in the 
External Temperature
 box.
c.
Click 
Apply
.
d. Ensure that neither polled variable is configured to poll for temperature. Polling for 
pressure is allowed. See Section 5.8.