Emerson Network Router Manuel D’Utilisation

Page de 110
Installation and Use Manual
73
Meter Verification
Met
e
r V
e
rification
T
ransmitter T
e
rmina
ls
Data Log
g
e
r
T
ra
n
sm
itte
r Co
nf
igu
rati
o
n
6.2.4
Fourth panel: Test Results
The Test Results panel, shown in Figure 6-4, summarizes the result of the meter verification test.
Figure 6-4
Meter Verification - Test Results panel
The indicator located at the top of the panel indicates the test result. It is green if the meter passed the 
test and red if it failed.
In addition to the pass/fail indicator, ProLink II shows the changes in tube stiffness on a graph. This 
allows you to see not only whether the meter is operating within specification, but also where the 
results fall within the specified limits. For each test, the results are shown as two data points 
corresponding to the stiffness at the inlet and at the outlet of the flow tubes.
This graph also shows the results of previous tests performed on this meter. The rightmost data points 
are the most recent. This history lets you see how your meter is trending over time, which can be a 
useful way to detect meter problems before they become severe. The trending of both the inlet and the 
outlet points over several tests can also help determine whether changes to the flow tubes are localized 
or generalized.
Note: Historical data are stored on the computer on which ProLink II is installed. If you performed 
meter verification tests on the same meter from a different computer or from the meter display, the 
historical data for these tests will not be visible.