Honeywell SC1000 Manuel D’Utilisation

Page de 174
 page 82      008-0608-00
11.4  Specifications
 TRANSDUCER INPUT
Transducer type
full-bridge, AC/AC-LVDT 
Excitation Voltage
3 VRMS @ 5kHz,
short circuit protected
Transducer full-scale output
.1 to 15 VRMS @ 3 VRMS excitation
Amplifier Gain Selection
automatic 
Calibration Type
2-, 3- or 5- point known displacement
A/D Converter
24-bit Sigma-Delta
Low-pass filter
digital, 24-tap FIR
Resolution
Frequency Response
AUXILLIARY INPUTS
Quantity
2
Type
momentary contact closure
Response Time
< 5ms
Field-Selectable Functions
tare on, tare off,
peak/valley clear, peak/valley hold,
track hold
ANALOG OUTPUT
Voltage range  (voltage output
channels)
5, ±5, 10 or ±10 VDC
Current range (current output
channels)
4-20 mA
Source
any channel’s track, peak or valley value
Isolation
500V
Resolution
13 bits
Frequency Response
same as input if driven by the same chan-
nel’s tracking data