Chapter 13
13-4
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
Attaching the Sensor 
to the Ventilator
The O
2
 sensor attaches to the gas outlet port located below the front panel on 
the ventilator’s lower right corner.
Figure 13-2: O
2
 Sensor Orientation
1. Rotate the sensor assembly so that the sensor is pointing up. 
Figure 13-3: Connecting O
2
 Sensor and Patient Circuit
2. Connect the T-fitting (2) to the gas outlet port (1).
3. Insert the inspiratory bacteria filter (4) into the T-fitting (2) using the 
22mm connector (3) if necessary. Some bacteria filters provide an 
arrow or other mark to indicate the direction of flow. The flow indicator 
should be pointed away from the ventilator, toward the patient circuit 
connection.
CAUTION:
Flow
1
2
3
4
4
3
2
1
5
5