Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-5
Chapter 13
4. Connect inspiratory limb of patient circuit (5) to bacteria filter (4).
5. Ensure that all connections are tight. 
Warranty
Respironics warrants the O
2
 sensor to be free from defects in material and 
workmanship for a period of one year from the date of purchase, provided that 
the unit is operated under conditions of normal use as described in this 
operator’s manual.
At its discretion, Respironics will make replacements, repairs, or issue credits 
for equipment or parts that are found to be defective.
O
2
 Sensor Tee
The External Oxygen Sensor Adapters (P/Ns 8-100498 and 1020380) are also 
referred to as O
2
 Sensor Tees. The PVC adapter is gray, the Ultem® is clear 
with an amber tint, and both are compatible with the MSA and Analytical 
Industries External Oxygen Sensors. Both tees are compatible with all 
ventilators.
Both tees have 22mm female/male connections; therefore the 22mm male-
male adapter (P/N 1002505) is not needed when using this adapter.
Figure 13-4: Sensor Tee
NOTE:
Record O
2
 sensor manufacturing or warranty numbers and installation 
date for future reference.
NOTE:
To ensure accurate O
2
 monitoring, check O
2
 sensors periodically and 
replace as per manufacturer specification.
NOTE:
Sensor performance and expected operating life information is outlined 
in the sensor manufacturer’s instructions for use. Thoroughly review all 
O
2
 sensor instructions prior to installation and use with the ventilator.
NOTE:
O
2
 sensor calibration is performed during EST. If recalibration of the O
2
 
sensor is required, follow the instructions for “Extended Self Test 
(EST)” in Chapter 11.
CAUTION:
8-100498-00
1020380
Ultem®
PVC
(Gray)
(
Clear)