Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-11
Chapter 13
Testing
The external battery should be tested during the regular preventative 
maintenance cycle of the ventilator or as required by your institutional 
procedures. Perform the External Battery Verification Procedure included in 
the External Battery Service Instructions (1013569).
Battery 
Specifications
*Batteries are considered hazardous waste and must be disposed of according 
to local regulations. It is unlawful to dispose of these batteries except through 
a regionally approved recycling center.
For additional information, contact Respironics Customer Service at 
1-800-345-6443.
CAUTION:
The External Battery is designed to recharge only as described here. Do 
not use any other method to recharge the External Battery.
CAUTION:
If the ventilator is expected to be stored for more than 2 weeks, the 
external battery ON/OFF switch should be set to the OFF (O) position to 
avoid discharging the external battery. This switch must be placed in 
the ON (I) position when the ventilator is removed from storage and 
returned to use.
Part Number:
1059955
Model:
V200 External Battery
Type:
Lead Acid*
Output:
24 V, 33AH
Class:
1
Recharge Time:
~ 15 hours
Charge Hold:
~ 14 days at 20 C
Storage Conditions:
-20 C to 60 C at 10%-95% relative humidity