Battery Capacity
The external battery is used with the backup battery. Under optimal running 
conditions, both batteries together can operate the ventilator for up to four 
hours. However, ventilator power consumption varies according to 
environmental conditions and ventilator settings. Battery capacity varies with 
the age and charge level of the battery. The external battery should be kept 
fully charged to ensure maximum capacity when needed.
Battery Charging
The external battery is designed to be recharged only as described here. Do not 
use any other method to recharge the battery.
In order for the external battery to charge, the system must be connected to a 
viable AC power source. The ventilator can be operating or in standby mode. A 
new external battery fully charges within 15 hours. The time required to 
recharge depends on the original charge level of the battery.
When the external battery is connected to a functioning AC power source, the 
MAINS 
indicator located on the front of the ventilator is lit and the green AC 
indicator on the external battery is lit (Figure 13-7).
The yellow “Battery Charging” indicator on the external battery has the 
following indication modes:
Continuous ON - Battery is in deep charge mode
Flashing - Battery is approaching full charge
Off-Battery is fully charged
Figure 13-7: External Battery Indicator Lights
If the external battery does not reach full charge within 15 hours, contact 
Respironics Customer Service at 1-800-345-6443.
CAUTION:
Battery operating life depends on battery age and number of discharges 
and recharges. Over time the battery degenerates and provides less 
operating time per charge than a fully-charged new battery. Use a 
Respironics External Battery P/N 1059955
 
only.