Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-57
Chapter 13
3. At the end of a successful maneuver, a pop-up screen displays the 
result of the maneuver and asks you to accept or reject the results by 
touching the Accept or Reject button.
4. At the end of the maneuver (whether successful or cancelled), the 
Stop button changes to Continue. While the Continue button is 
displayed, the graph is frozen and can be rescaled. Touch Continue or 
exit the screen to unfreeze the display.
5. Touch Continue: the button changes to Start and you can repeat the 
maneuver. 
6. Return the active ventilation mode from VCV to the previous mode if 
necessary.
Figure 13-31: Static C and R Screen
Information type
Description
Pressure--Volume or 
Pressure--Time button
Pressure--Volume: Displays a real-time pressure-volume loop.
Pressure--Time: Displays a real-time pressure-time waveform.
Start button
Starts the maneuver. Changes to Stop when the maneuver is started. 
Changes to Continue when the maneuver is complete or when the Stop 
button is touched. Changes to Start when Continue is touched.
Static C and Static R 
windows
Displays the most recently accepted maneuver results. The date and 
time of the maneuver appear below each window.
Table 13-7: Summary of Static C and R Screen Information
Time 
Stamp
Rescale
button
Start
button
Static C and R Windows
Pressure-Volume and Pressure-Time buttons
P Plat 
window
Scale 
button