Chapter 13
13-58
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
Alarms and Error 
Messages
If an alarm occurs during Respiratory Mechanics, an Alerts window lists the 
active alarms. Any active maneuvers are cancelled if a high-priority alarm 
occurs. If there is a preexisting alarm of any priority when you attempt to start 
a maneuver, the maneuver cannot be performed and an error message is 
displayed. Table 13-8: “Respiratory Mechanics Error Messages and Alarms” 
summarizes error messages and changes that can occur during Respiratory 
Mechanics.
P
plat window
Displays the most recently accepted maneuver results. The date and 
time of the maneuver appear below each window.
Scale (button at left of 
graphic display)
Allows you to adjust graphic scales manually. The default scales for the 
Static C & R screen are 0 cmH
2
O to the High Pressure limit for 
pressure, 0 to Vt + 100 mL for volume. Default scales are in effect 
every time you enter the Static C and R screen.
Rescale button
Allows software to adjust the vertical scales for waveforms and vertical 
and horizontal scales for loops.
Information type
Description
Table 13-7: Summary of Static C and R Screen Information