Philips V200 Manuel D’Utilisation

Page de 328
Chapter 8
8-6
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Operating Instructions
Power Status Indicators
Power status indicators alert you to the status of the backup battery. The text 
version of the power status indicators is shown in Figure 8-5. The symbol 
version of the power status indicators is shown in Figure 8-6.
NOTE:
Figure 8-5: Power Status Indicators — Text Version
Figure 8-6: Power Status Indicators — Symbol Version
Power Status Indicators 
Battery Indicator
Status
Description
In Use
Yellow (active)
The ventilator is running on backup battery 
power. The backup battery is used when there 
is no AC power and no other battery is 
available.
Charging
Yellow (active)
The backup battery is charging. The battery 
should not be considered a fully charged power 
source when this indicator is illuminated. The 
Charging indicator will stay on for the duration 
of the charging cycle, which can last up to ten 
hours. Once the battery has fully charged, the 
Charging indicator will turn off.
Low
Flashing red
The backup battery has approximately 5 
minutes of power remaining.
Table 8-2: Power Status Indicators  (Sheet 1 of 2)