5 - 18
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Calibration parameters
↓↓
Advanced calibration methods
↓↓
In the menu "Advanced Calibration Methods" you can
♦ 
start the zero calibration of all measurement ranges for all channels
of an MLT analyzer or analyzer module simultaneously or
♦ 
start the span calibration of all measurement ranges for all channels
of an MLT analyzer or analyzer module simultaneously or
♦ 
start the zero and span calibration together for the current channel
of an MLT analyzer or analyzer module or
♦ 
start the zero and span calibration together for all channels
of an MLT analyzer or analyzer module or
♦ 
set all calibration deviations to zero ("basic calibration procedure").
Starting the functions:
♦ 
Change to the line you want with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Press the 
88 
 -key or the 
 -key or
cancel and go back to the last menu page with the F4 key (Back...) or the 
 -key.
♦ 
If asked: Confirm with the F2 key (Yes) or
cancel and go back to the last menu page with the F4 key (Back...) or the 
 -key.
Measure
Status...
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Advanced Calibration Methods --
Start zero calibration procedure for all channels !
Start zero+span calibration procedure !
Start zero+span calibration procedure for all channels !
Start basic calibration procedure !
Info...
Start span calibration procedure for all channels !
Cancel all running procedures !
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Advanced calibration methods